Method 2


Cleaning Method 2

Product Care Instructions

For All: Ocular MaxField® (Glass) and MaxLight® (CR-39) Indirect Diagnostic/Laser Lenses

Except: (See Cleaning Method 3)
OI-20A Ocular MaxACTM 20 Diopter
OI-28A Ocular MaxACTM 28 Diopter

Cleaning

Wipe: Clean with alcohol wipe.
Then: Proceed with either disinfection or sterilization instructions.

Disinfection
Soak In: GLUTARALDEHYDE 2% or 3.4% aqueous solution
Temperature per manufacturer instructions
Minimum exposure time = 20 minutes or

BLEACH 10% solution mixed at: 1 part bleach to 9 parts water
Recommended exposure time = 10 minutes

CAUTION To avoid damage to the lens, do not exceed recommended exposure time.

Then: Rinse lens thoroughly to remove disinfection solution.
3 cycles of 1 minute, with cool or tepid water is recommended.

Dry carefully and place in a dry storage case.

NOTE These lenses are known to be compatible with:
Ascepti-Wipe, Cavi-cide, Cidex, Cidex OPA, DisCide Wipe, Enviro-cide, and Opti-Cide.

MaxField® lenses are also compatible with H2O2 -3%.

Sterilization - EO

Minimum Time: 1 hour
Temperature: 130°F (54°C)
Aeration Time: 12 hours

Sterilization - Steam Autoclave

No

Sterilization - Steris System 1

Follow manufacturer's instructions for MaxField® (glass) lenses.
Not compatible with MaxLight® (CR-39) lenses.

Sterilization - Sterrad

No

WARNING Never steam autoclave or boil listed lenses.
Never soak in Alcohol, Acetone or Other Solvents.

For information on compatibility with alternative product care methods, contact Customer Service.

For information on compatibility with alternative product care methods, contact Customer Service.

Tel: 1-800-888-6616 (USA)
Email: contact@ocular-instruments.com

Cleaning Method 2
Alternate Language Instructions
(PDF download)


METODO DI PULIZIA 2
MÉTHODE DE NETTOYAGE 2
REINIGUNGSMETHODE 2
METODO 2 DE LIMPIEZA