Method 1


Product Care Instructions
For All: Ocular Argon/Diode, YAG and Diagnostic Lenses

Plus: OKSG, Khaw Surgical Gonioprism
OLIV-EQNA, Landers NA Equatorial Vitrectomy Lens
OLIV-WFNA, Landers NA Wide Field Vitrectomy Lens
OLTK-7.2 or 8.2, Landers Wide Field Temporary Keratoprosthesis
OTSG, Thorpe Surgical Gonioscope
OWIV-HMNA, Woldoff NA High Mag Vitrectomy Lens
OUV 132-2, Peyman-Wessels-Landers 132D Upright Vitrectomy Lens

Except: (See Cleaning Method 3)
O4MAC Ocular MaxField® AC Four Mirror Gonio Lens
O4MAC-LR Ocular MaxField® AC Four Mirror Gonio Lens
O4MAC-H Ocular MaxField® AC Four Mirror Gonio Lens
OG3MAC-10 Ocular Autoclavable Three Mirror Gonio 10mm

Cleaning

Rinse: Immediately upon removal from patient's eye, thoroughly rinse in cool or tepid water.
Wash: Place a few drops of mild soap on a moistened cotton ball. Gently clean with a circular motion.
Rinse: Thoroughly rinse in cool or tepid water, then dry carefully with a non-linting tissue.
Then: Proceed with either disinfection or sterilization instructions.

Disinfection

Soak In: GLUTARALDEHYDE 2% or 3.4% aqueous solution Temperature per manufacturer instructions
Minimum exposure time = 20 minutes or

BLEACH 10% solution mixed at: 1 part bleach to 9 parts water
Recommended exposure time = 10 minutes

CAUTION
To avoid damage to the lens, do not exceed recommended exposure time.

Then: Rinse lens thoroughly to remove disinfection solution.
3 cycles of 1 minute, with cool or tepid water is recommended.
Dry carefully and place in a dry storage case.

NOTE These lenses are known to be compatible with:
Ascepti-Wipe, Cavi-cide, Cidex, Cidex OPA, DisCide Wipe, Enviro-cide, and Opti-Cide.

Also compatible with H2O2 -3%, except the following lenses:
OG3M-10, Three Mirror 10mm Diagnostic
OPDSG, OPDSG-2, OPDSG-3, Posner Gonioprisms
OS4M, OS4M-2, Sussman Gonioscopes
OK4DG, Khaw Direct View Gonio

CAUTION If used on an ulcerated cornea, lens must be STERILIZED before next procedure.

Sterilization - EO

Minimum Time: 1 hour
Temperature: 130°F (54°C)
Aeration Time: 12 hours

Sterilization - Steam Autoclave
No
Sterilization - Steris System 1
Follow manufacturer's instructions.

Steris is not compatible with Posners (OPDSG, OPDSG-2, OPDSG-3), Sussmans (OS4M, OS4M-2), OK4DG, Mainsters (OMRA-S, OMRA-S-2, OMRA-HM, OMRA-HM-2, OMRA-WF, OMRA-WF-2), OG3M-10, OGP2, OGP2H

Sterilization - Sterrad
No
WARNING Never steam autoclave or boil listed lenses.
Never soak in Alcohol, Acetone or Other Solvents.

For information on compatibility with alternative product care methods, contact Customer Service.

Tel: 1-800-888-6616 (USA)
Email: contact@ocular-instruments.com

Cleaning Method 1